EMC seminar
Transient Limiter
The purpose of the limiter is to protect the input of the EMC analyzer from
large transients when connected to a LISN. Switching EUT power on or off
can cause large spikes generated in the LISN.
The Agilent 11947A transient limiter incorporates a limiter, high-pass filter,
and an attenuator. It can withstand 10 kW for 10 μsec and has a frequency
range of 9 kHz to 200 MHz. The high-pass filter reduces the line frequencies
coupled to the EMC analyzer.
Page 35
DUT
Limiter
LISN